IEEE-488 Reference
5-72
:ENABle <b>
:DISPlay:ENABle <b> Control display circuitry
<b> = 0 or OFF Disable display circuitry
= 1 or ON Enable display circuitry
:disp:enab <b>
Power-up ON
*RST No effect
:SYSTem:PRESet No effect
:ENABle? Query state of display
Short-form formats: :disp:enab?
Response message: 1 (on) or 0 (off)
This command is used to enable and disable the front panel display circuitry. Disabling the dis-
play circuitry allows the instrument to operate at a higher speed. While disabled, the display will
be frozen with the following message:
FRONT PANEL DISABLED
Press LOCAL to resume.
As reported by the message, all front panel controls (except LOCAL) will be disabled. Normal
display operation can be resumed by using the :ENABle command to enable the display or by
putting the Model 7001 into local.
10 OUTPUT 707; “:disp:enab off; enab?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Disables the display, and then queries the state of the display.
Line 20 Addresses the 7001 to talk.
Line 30 Displays (on the CRT) the programmed state of the display (0; off).
Parameters
Short-form
formats
Defaults
Query
command
Description
Programming
example
:DISPlay subsystem