Front Panel Operation
4-50
#2. When the measurement is completed, a trigger is
sent to the Model 7001 to close the next channel. After
this channel settles, a trigger is sent to the DMM to trig-
ger a measurement. This process continues until all 10
channels are scanned and measured. The data store of
the Model 196 could be used to store the measurements
as they occur.
Figure 4-16
DUT test system
196 DMM
1
DUT
#1
2
DUT
#2
10
DUT
#10
OUTPUT
Card 1
7011 MUX Card
HI
LO
Figure 4-17
External trigger connectors
Channel
Ready
External
Trigger
7001 Switch System
7051-2
BNC to BNC
Cables (2)
External
Trigger
Input
Voltmeter
Complete
Output
196 DMM
Ready of the Model 7001 to External Trigger Input of
the Model 196 to trigger a measurement of DUT #1. Af-
ter the Model 196 completes the measurement, it will
output a trigger from Voltmeter Complete Output to
External Trigger input of the Model 7001, which closes
the next channel. After Channel 1!2 settles, a trigger is
sent to the Model 196 to trigger a measurement of DUT