Chapter 7 Assembly of the Devices
151
7.10 Attaching an Analyzer and a Polarizer (Required for DIC
Microscopy)
7.10.1 Attaching an Analyzer
Attach an analyzer to the microscope main body.
Alternatively, an analyzer cube can be attached to the FL turret.
The attachment procedure for analyzer cubes is the same for fluorescent filter cubes. For details, see “7.5.2
Mounting a Filter Cube.”
1. Remove the dust-proof slider from the
microscope main body.
2. Insert the TI2-C-DICA analyzer.
The analyzer slider can be inserted from both
sides of the microscope main body.
Setting the slider to second click position will
place the analyzer into the optical path.
Returning to the first click position will remove
the analyzer from the optical path, with a hollow
hole placed in the optical path instead.
A
nalyzer slide
r
A
nalyzer slo
t