EasyManuals Logo
Home>Nikon>Microscope>ECLIPSE Ti2-U

Nikon ECLIPSE Ti2-U User Manual

Nikon ECLIPSE Ti2-U
203 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
Chapter 3 Usage of Components
66
3.10.4 Using an Objective with a Diaphragm
When performing DF microscopy, use of an objective
with diaphragm allows numerical aperture (NA)
adjustment of the objective according to the NA of
the DF condenser.
Operation:
Check the minimum NA of the DF condenser, and
then adjust the diaphragm adjusting ring so that
the NA of the objective is smaller than the
minimum NA.
Display:
Check the numerical aperture using the gradua-
tions on the diaphragm adjusting ring.
Example of an objective with a diaphragm
“Iris” mark
Diaphragm
adjusting ring

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon ECLIPSE Ti2-U and is the answer not in the manual?

Nikon ECLIPSE Ti2-U Specifications

General IconGeneral
BrandNikon
ModelECLIPSE Ti2-U
CategoryMicroscope
LanguageEnglish

Related product manuals