Chapter 8 Specifications
183
8.3 Physical Properties
Model ECLIPSE Ti2-U
Electrical shock
protection class
Class I
Class I (AC adapter)
Environmental conditions Operation: Temperature: 0 to 40°C
Humidity: 60% RH max. (at +40°C, no condensation)
Storage and Transport: Temperature: -20 to +60°C
Humidity: 90% RH max. (no condensation)
Altitude: 2,000 m max.
Pollution: Degree 2
Overvoltage category: Category II
Indoor use only
External dimensions and
mass
Ti2-U microscope system
(Device combinations: Ti2-U, TI2-D-PD, TI2-D-LHLED, TI2-C-TC, TI2-T-BC,
TC-T-ER, TC-S-SR, TI2-N-ND-I, TI2-F-FLT-I, TI2-LA-BF, TI2-LA-FL-2)
External dimensions: 390 (W) x 668 (H) x 626 (D) mm (excluding protruding
parts)
Mass (reference): Approximately 37.6 kg
Ti2-U microscope main body
External dimensions: 240 (W) x 378 (H) x 527.5 (D) mm (excluding
protruding parts)
Mass (reference): Approximately 20 kg
Safety standards
Configuration of microscope:
• Ti2-U
• TI2-D-PD
• TI2-D-LHLED
• TI2-C-DICP
• TC-C-TC
• TI2-T-BC
• TI2-N-ND
• TI2-F-FLT-I
• TI2-LA-BF
• TI2-LA-FL-2
• TI2-LA-SU
• CE Marking
• IVD Directive
This equipment complies with the emission and immunity
requirements of IEC/EN 61326-2-6.
• Low Voltage Directive
• EMC Directive
• C-UL-US Listed
• FCC Part 15 Subpart B Class A
Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful inter-
ference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful inter-
ference in which case the user will be required to correct the interference at the
user's expense.
• CAN ICES-003(A) / NMB-003(A)
• Australian EMI (AS/NZS CISPR11)