Chapter 8 Specifications
180
Diascopic illuminator
TI2-D-PD
Pillar for Dia Illumination
Condenser holder
Fixed filter slot
Illumination method: Köhler illumination
Magnification of the light source: LED1.3x, 1.7x (used with a high NA lens)
Halogen lamp: 4.5x, 5.8x (used with a high NA lens)
Field diaphragm: Diameter of 34 or more to diameter of 1.2 or less
Shutter: Manual
Condenser vertical stroke: 66 mm, with the refocusing mechanism
Condenser fixing section: Round dovetail mount, with the centering mechanism
Polarizer fixing section: Diameter 63.8 mm, joint
Number of slots: 2-Position, fixed
Supported filters: Diameter 45 mm, thickness 3 mm or less
TI2-D-LHLED
LED Lamp House for Dia
Illumination
LED used: White LED
Power source for illumination: Supplied from the Ti2-U main body
TI2-D-SF
Filter Slider for Dia
Illumination
Holder type: Two-hole filter slider (x 2), slide system
Fixed filter slot: Two positions
Supported filters: Diameter 45 mm, thickness 3 mm or less
D-LH/LC
Precentered Lamphouse
Input rating: 12 VDC 100 W
Lamp rating: 12 V 100 W Halogen lamp
Designated lamp: Halogen lamp
(OSRAM HLX 64623 or PHILIPS 7724I)
Condenser
TC-C-TC
Condenser turret
Number of modules: 7
Module mount holes: Small opening (diameter 37 mm), 4 (for Ph, NAMC,
and ND)
Large opening (diameter 39 mm), 3 (DIC, IMSI, and
ND)
Turret drive method: Manual
Detection of turret position signals: None
Lens mount: M48 x 0.75
Supported condenser lenses: LWD, ELWD, CLWD, and NAMC condensers
Aperture diaphragm: Diameter 34 to diameter 2 or less
Shutter: Shutter addresses using a shielding plate can be set.
TI2-C-SCH
HNA condenser slider
Number of modules: 2
Module mount holes: Small opening (diameter 37 mm), 1 (for Ph, and ND)
Large opening (diameter 39 mm), 1 (for DIC and ND)
Slider drive method: Manual
Lens mount: M48 x 0.75
Supported condenser lenses: HNA dry lens, HNA oil lens
Aperture diaphragm: Diameter 34 to diameter 2
TE-C
ELWD-S Condenser
Condenser lenses: NA0.3, OD65
Annular diaphragm: PhL, Ph1, Ph2
Lens mount: M48 x 0.75
Aperture diaphragm: Diameter 27 to diameter 2 or less