EasyManuals Logo

Olympus ESG-410 User Manual

Olympus ESG-410
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
32
Installation
4.5 Connection of foot switches
Up to 3 foot switches can be connected to the rear panel of the electrosurgical generator.
However, for safety reasons it is not possible to use foot switches of the same type
simultaneously, e.g. 2 single-pedal foot switches or 2 wireless double-pedal foot switches.
Therefore, ensure that only different types of foot switches are connected, e.g. 1 single-
pedal foot switch, 1 wireless double-pedal foot switch and 1 cabled double-pedal foot switch.
For detailed information on correctly connecting foot switches, refer to the instructions for
use of the compatible foot switches.
For information on compatible foot switches refer to the chapter “15 Compatible equipment”
on page 95.
4.6 Connection of peripheral equipment
Peripheral equipment can be connected to the LINK-IN socket and the LINK-OUT socket
on the rear panel of the ESG-410. When connecting peripheral equipment, consider the
following:
The peripheral equipment must comply with respective IEC or ISO standards, e.g.
IEC 60950-1/62368-1 for data processing equipment or IEC 60601-1 for medical
equipment.
For the connection the compatible communication cables “MAJ-1871” (0.25 m) or
“MAJ-1872” (10 m) must be used.
Also refer to the peripheral equipment’s instruction for use.
Both devices, the ESG-410 and the peripheral equipment, must be switched off before
connecting.
4.7 Connection for automatic smoke evacuation
When the electrosurgical generator is combined with the Olympus high ow insufation
unit UHI-2, UHI-3 or UHI-4, smoke and mist can be evacuated simultaneously with the
electrosurgical output.
4.7.1 Connecting the ESG‑410 to UHI‑2 or UHI‑3
When connecting the ESG-410 to the UHI-2/3 also refer to the instructions for use of the
UHI-2/3.
Preconditions
UHI-2 or UHI-3 is available
Communication cable “MAJ-1872” (10 m) is available
Adapter “MAJ-1873” is available
ESG-410 and UHI-2/3 are switched off

Table of Contents

Other manuals for Olympus ESG-410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus ESG-410 and is the answer not in the manual?

Olympus ESG-410 Specifications

General IconGeneral
BrandOlympus
ModelESG-410
CategoryPortable Generator
LanguageEnglish

Related product manuals