62
Before use
Maximum output current .................................................................................................. 0.8 A
Rated load ....................................................................................................................... 500 Ω
7.5 Tissue effects depending on the power level and effect level
7.5.1 Bipolar modes
Mode name Results when increasing/decreasing
the effect level
Results when increasing/decreasing
the power level
BipolarCut
PlasmaCut
PK LoopCut
PK MorceCut
PK PureCut
PK SoftCut
Increased / decreased thermal effect Increased / decreased thermal effect and
increased / decreased cutting capability
AutoCoag
BiSoftCoag
PK Coag
PK SoftCoag
Increased / decreased coagulation speed Increased / decreased coagulation depth
for short time application.
Nevertheless, to achieve the full potential
coagulation depth a long application time
with a low power setting is necessary.
HardCoag Increased / decreased automatic end of
procedure detection time
Increased / decreased thermal effect
SalineCoag Increased / decreased thermal effect Increased / decreased thermal effect
7.5.2 Monopolar modes
Mode name Results when increasing/decreasing
the effect level
Results when increasing/decreasing
the power level
Blend
BlendCut
PulseCut Fast
PulseCut Slow
Pure
PureCut
Increased / decreased thermal effect Increased / decreased thermal effect and
increased / decreased cutting capability
FineCut n/a Increased / decreased thermal effect and
increased / decreased cutting capability
PowerCut Increased / decreased dissection /
coagulation
Increased / decreased thermal effect
Fulgurate
PowerCoag
Increased / decreased coagulation /
dissection capability
Increased / decreased thermal effect
ForcedCoag
Spray
SprayCoag
Increased / decreased coagulation speed Increased / decreased thermal effect