41
System settings
5.10 Startup settings
[Previous settings] and [All zero]
The ESG-410 provides 2 options on how to deal with HF output settings when the device is
switched off:
[Previous settings]
The selected HF output settings are stored. When the ESG-410 is switched on again,
these settings are automatically displayed.
[All zero]
The selected HF output and settings are deleted. When switching on the ESG-410
again, all HF output settings are automatically set to [Off].
The default setting is [All zero].
Selecting the settings for startup
1. Tap [Settings] on the home screen.
The [Setup] screen is displayed.
2. Tap the options button for [Startup settings].
The [Startup settings] screen is displayed. The selected setting is highlighted.
3. Tap the button for the preferred setting.
4. Tap [Back] to return to the [Settings] screen.
The updated setting is displayed.
[Last settings] button
If the ESG-410 is switched off unintentionally, then the last HF output settings can be
restored by tapping this button.
If [All zero] was chosen as a startup setting, then this button is displayed on the top of the
home screen when turning on the ESG-410.
The [Last settings] button is also available on the [Startup settings] screen.
5.11 Procedures
Use this function to save the current settings as a user-dened procedure or to select a user-
dened procedure from a list.
Data that are saved within a user-dened procedure are:
All selected modes
All selected power settings
All selected effect setting
All selected level settings
Instrument names
The foot switch assignments
The Autostart assignment
The Autostart delay
The setting for the baby neutral electrode
The selection if an output socket is switched on or off.
User-dened procedures can be stored on a portable USB ash drive. Refer to the section
“11.2.1 Data transfer” on page 87.
5.11.1 Saving a user-dened procedure
1. Select all settings on the touchscreen that are required for the new procedure.

Table of Contents

Other manuals for Olympus ESG-410

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus ESG-410 and is the answer not in the manual?

Olympus ESG-410 Specifications

General IconGeneral
BrandOlympus
ModelESG-410
CategoryPortable Generator
LanguageEnglish

Related product manuals