99
Technical data
16 Technical data
16.1 Specications for the ESG-410
Size
Width ......................................................................................................................370 ± 2 mm
Height ..................................................................................................................... 241 ± 2 mm
Depth ......................................................................................................................501 ± 2 mm
Depth including connected cables .......................................................................520 ± 10 mm
Weight ........................................................................................................................... 16.4 kg
Weight of packaging ...................................................................................................... 2.48 kg
Power supply
Voltage range ...................................................................... AC 100 to 120 V/AC 220 to 240 V
Frequency ............................................................................................................ 50 Hz/60 Hz
Maximum input power for AC 100 to 120 V ................................................................ 1150 VA
Maximum input power for AC 220 to 240 V ................................................................ 1500 VA
Fuse ............................................................................................................ 2 x 250 V/T 10 AH
Classications
According to IEC 60601-1, protection against electric shock:
Medical electrical equipment ..........................................................................................Class I
Applied parts .............................................................................................................. Type CF
Protection against harmful ingress of water ...................................................................... IP20
Mode of operation ............................................................................Non-continuous operation
According to applicable European Union harmonization legislation .....................................IIb
Output
High frequency functions....................................................................... Monopolar and bipolar
High frequency at rated load .............................................................................344 to 516 kHz
Maximum HF power monopolar .......................................................................300 W at 500 Ω
Maximum HF power bipolar ...............................................................................320 W at 75 Ω
All modes ...................................25% duty cycle (e.g. activated for 10 s, deactivated for 30 s)
16.2 Ambient conditions for operation, storage and transport
Temperature
For operation and storage .................................................................................. +10 to +40 °C
For transport........................................................................................................-25 to +60 °C
Relative humidity
For operation and storage ............................................................ 30 to 85 %, non-condensing
For transport................................................................................. 10 to 85 %, non-condensing
Atmospheric pressure
For operation ................................................................................ 700 to 1060 hPa (≤ 3000 m)
For storage and transport ............................................................. 500 to 1060 hPa (≤ 3000 m)