9-14 Operator’s Manual – OPTI CCA-TS2
9 OPERATING PRINCIPLES
9.5.11 Accessories
OPTISensorCassettes
Type“B”,BP7562(pH,PCO
2
,PO
2
,tHb,SO
2
)
Type“E”,BP7587(pH,PCO
2
,PO
2
,Na
+
,K
+
,tHb,SO
2
)
Type“E-Ca”,BP7560(pH,PCO
2
,PO
2
,Na
+
,K
+
,Ca
++
,tHb,SO
2
)
Type“E-Cl”,BP7559(pH,PCO
2
,PO
2
,Na
+
,K
+
,Cl
-
,tHb,SO
2
)
Type“E-Glu”,BP7564(pH,PCO
2
,PO
2
,Na
+
,K
+
,Glu,tHb,SO
2
)
Type“E-BUN(urea)”,BP7588(pH,PCO
2
,PO
2
,Na
+
,K
+
,BUN(urea),tHb,SO
2
)
Type“B-Lac”,BP7561(pH,PCO
2
,PO
2
,Lac,tHb,SO
2
)
Type“B60”,BP7586(pH,PCO
2
,PO
2
)
Use: For measurement of various analytes with the OPTI CCA-TS2 Analyzer.
Contents: Box contains 25 individually packaged cassettes. Each disposable
plastic cassette contains buffer and optical sensors.
Composition: Aqueous HEPES-bicarbonate buffer solution 0.2 mL with biocides.
Storage: Refer to package labeling.
Stability: Expiration date and lot number are printed on each cassette container
label.
Multi-LevelStandardReferenceCassettes(SRCs)BP7652
Use: For diagnostic and daily QC check of the OPTI CCA-TS2
Contents: Each package contains one reusable SRC Cassette.
Composition: Stabilized optode sensors with assay values:
Level1 Level2 Level3
pH 7.080 - 7.120 7.380 - 7.420 7.580 -7.620 pH units
PCO
2
68.0 - 72.0 38.0 - 42.0 18.0 - 22.0 mmHg
PO
2
57.0 - 63.0 97.0 - 103.0 167.0 - 173.0 mmHg
Na
+
123.0 - 127.0 143.0 - 147.0 163.0 - 167.0 mmol/L
K
+
2.2 - 2.8 4.2 - 4.8 6.7 - 7.3 mmol/L
Ca
++
1.7 - 1.9 1.0 - 1.2 0.6 - 0.8 mmol/L
Cl
-
78.0 - 82.0 103.0 - 107.0 128.0 - 132.0 mmol/L
Glu 36.0 - 44.0 106.0 - 114.0 296.0 - 304.0 mg/dL
Glu 2.00 - 2.44 5.88 - 6.33 16.43 - 16.87 mmol/L
BUN 4.2 - 7.0 26.6 - 29.4 68.6 - 71.4 mg/dL
Urea 1.5 - 2.5 9.5 - 10.5 24.5 - 25.5 mmol/L
Lac 0.70 - 1.30 2.00 - 3.00 4.50 - 5.50 mmol/L
Lac 6.3 - 11.7 18.0 - 27.0 40.5 - 49.5 mg/dL
tHb 18.5 - 21.5 12.5 - 15.5 6.5 - 9.5 g/dL
SO
2
68.0 - 72.0 88.0 - 92.0 96.0 - 100.0 %
Storage: Refer to package labeling.
Stability: Expiration date and lot number are printed on each package label and
encoded on the attached bar code label.