3-32 Operator’s Manual – OPTI CCA-TS2
3 CUSTOMIZATION
To adjust the tracking barometer within the OPTI
CCA-TS2, follow the instructions below:
1. In the main menu, select<System Manager>
and <Setup>.
2. In the <System Setup> menu, press the
<System> tab and then <Hardware> to select
this function (Fig. 3-73).
3. Press to enter an offset from the true
barometric pressure (Fig. 3-74).
4. Type in the new numbers and press to
accept the changes.
5. Press to return to the <Setup> screen or
to return to the main menu.
CAUTION:
Usetheabsolutebarometricpressure
andnotthealtitude-correctedpressure
(checkwithyourlocalweatherservice
orairport).
NOTE: Youmaychangebarometricpressureunits
frommmHgtombar(Seesection3.2.4.2).
NOTE: Youshouldcheckthebarometricpressure
periodically.
3.2.4.4 Hardware Settings
Fig. 3-73 Select Hardware
Fig. 3-74 Barometric pressure
The <Hardware> menu is used to adjust the local
barometric pressure, the audible alarm, and standby
mode.
3.2.4.4.1 EnteringtheBarometricPressure