Scale parameters and functions
8.4 Calibration
SIWAREX WP251
78 Manual, 12/2015, A5E37203357A
Image 8-4 Linearizing the scales' characteristic curve
L=0 100 kg load cell (2 mV/V) not
Approx. 0
L0 Mechanical installation on load
e.g. 25 kg
L1 Calibration weight 1 placed
e.g. 60 kg
L2 Calibration weight 2 placed
e.g. 80 kg Approx. 1 650 000
max
Nominal weight of load cell
Rated weight + approx. 10 %
Scales can be rapidly commissioned with automatic calibration. The accuracy of the scale
greatly depends on the entered parameters and the mechanical properties of the scale. The
best level of accuracy for the scale can be achieved by calibrating with calibration weights.
During initial commissioning with automatic calibration, the module must be reset using the
"Load factory settings" or "Load standard parameters" command.