Start and stop waveform acquisitions
From the Horiz/Acq menu, select Run/Stop.
Overview
Use the Run/Stop control window to start and stop waveform acquisition.
To use
â– 
Click Run/Stop to toggle waveform acquisition. (You can also press the front panel Run/Stop button.)
â– 
Click Single Sequence to complete a single acquisition sequence.
â– 
Click Force Trigger to immediately start acquiring waveform data without a valid trigger event.
Behavior
These controls operate identically to the front-panel controls.
â– 
Click Run/Stop once to stop waveform acquisition. The Status readout at the top of the graticule changes from Running to
Stopped and shows the number of acquisitions. Click Run/Stop again to resume waveform acquisition.
â– 
Click Single Sequence to provide a single sequence of acquisitions each time you click the button. When Single Sequence
is enabled, the front-panel single sequence (Single) button is lighted.
â– 
In Sample, Peak Detect, or Hi Res acquisition modes, the instrument acquires waveform records with the first valid trigger
event and then stops. In Average or Envelope acquisition modes, the instrument makes the specified number of acquisitions
to complete the averaging or enveloping task.
â– 
In Waveform Database mode, the instrument acquires and displays a waveform that is the accumulation of several
acquisitions. In addition to time and amplitude, the waveform displays a count of the number of times a specific sample point
has been acquired.
â– 
When you click Single Sequence (or press the Single button) in Equivalent Time mode, the instrument continues to
recognize trigger events and acquires samples until the waveform record is filled.
â– 
Click Force Trigger once to acquire a waveform record without a trigger event. Forcing a trigger is useful when the input
signal does not contain a valid trigger. The instrument recognizes and acts upon the forced trigger even when you click the
button before the end of the pretrigger holdoff. The button has no effect when acquisitions are stopped.
What do you want to do next?
Go to the Horizontal/Acquisition control window.
Return to the Acquisition Setup control window.
Return to the Trigger Setup control window.
Horizontal and acquisition setups
192 DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series

Table of Contents

Other manuals for Tektronix MSO5204B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO5204B and is the answer not in the manual?

Tektronix MSO5204B Specifications

General IconGeneral
BrandTektronix
ModelMSO5204B
CategoryTest Equipment
LanguageEnglish

Related product manuals