Behavior.
â–
Beep. When Beep is turned on, the instrument beeps when the test is completed. Beep is not available on MSO/DPO5000B
Series instruments.
â–
SRQ. When SRQ is turned on, the instrument sends a GPIB SRQ command when a test is completed.
â–
AUX Out. When AUX Out is turned on, the instrument sends a trigger to the Auxiliary Out connector when the test is
completed. See External Signals Control Window for additional details on this feature.
What do you want to do next?
Learn about mask polarity.
Learn about waveform parameters.
Go to the Test Setup control window.
Set mask test parameters
From the Mask menu, select Pass/Fail Setup.
Overview. Use the Number of Waveforms (Samples), Fail Threshold, and Pre-Test Delay controls to set the parameters for the
mask test.
NOTE. If you are using Waveform Database mode, the Number of Waveforms entry box changes to a Samples entry box.
To use.
1. Click in the Number of Waveforms (Samples) entry box to activate the multipurpose knobs to enter a value.
2. Click in the Fail Threshold entry box to activate the multipurpose knobs and enter the minimum number of waveforms that
must fail for the test to fail.
3. Click in the Pre-Test Delay entry box to activate the multipurpose knobs to enter the number of seconds to wait before
starting the test.
Mask testing setups