Behavior. The cursor units depend on the cursor type. The cursor position readouts appear on the display graticule. When using
the IRE graticule style, amplitude cursor readouts are shown in IRE units.
Cursor position does not change if the cursor source changes or if the waveform is repositioned.
TIP. Use the pop-up keypad or your external keyboard to specify exact numerical cursor coordinates.
What do you want to do next?
Learn about the Cursor Setup controls.
Learn about the cursor types.
Set display persistence
From the Display menu, highlight Display Persistence; the select Persistence Controls.
To use. Click one of the buttons to select the display persistence.
â–
Click Infinite (persistence) to continuously accumulate record points on the waveform until you change one of the acquisition
display settings.
â–
Click Variable; then click in the Time entry box and use the multipurpose knobs to set the persistence time.
â–
Click Reset to clear the persistence.
â–
Click Controls to activate the Persistence Control window.
â–
Click Setup to open the Display control window.
Behavior.
â–
Use infinite persistence for displaying record points that may occur outside the normal acquisition envelope.
â–
Use Variable to accumulate record points on the waveform for a specified time interval. Each record point decays
independently according to the time interval.
â–
If you click Off, the waveform record points appear for the current acquisition only.
TIP. Use the Display menu Display Persistence submenu or right-click in the graticule and select Display Persistence from
the shortcut menu to directly access the persistence options.
What do you want to do next?
Learn about changing the display colors.
Learn about using text with screen displays.
Learn about changing the appearance of objects in the display.