The following screen appears on MSO/DPO5000B Series instruments:
NOTE. To compensate for timing differences between voltage and current probes, order the following Probe Calibration and
Deskew Fixtures: Tektronix part number 067-0405-XX (for DPO7000C and MSO/DPO5000B oscilloscopes). Tektronix part
number 067-1586-XX (for MSO/DPO70000C/DX oscilloscopes). Tektronix part number 067-2083-XX (for MSO Series
instruments) to deskew the logic probe trigger path.
The fixtures may require a user-supplied pulse generator.
To use
1. Connect the probes to a signal source.
2. Use the horizontal and vertical controls to center the fastest signal on the screen.
3. Click a Chan <1–4> tab to select the slower channel.
4. Click in the Deskew Time entry box, and use a multipurpose knob to adjust the deskew time so that the two signals align.
5. To clear the deskew value, click the Set To 0.0 s button.
Vertical setups