14-40 IM 760401-01E
STATus?
Function Queries all settings related to the status of
communication.
Syntax STATus?
Example STATUS? -> :STATUS:EESE 0;
FILTER1 NEVER; FILTER2 NEVER;
FILTER3 NEVER;FILTER4 NEVER;
FILTER5 NEVER;FILTER6 NEVER;
FILTER7 NEVER;FILTER8 NEVER;
FILTER9 NEVER;FILTER10 NEVER;
FILTER11 NEVER;FILTER12 NEVER;
FILTER13 NEVER;FILTER14 NEVER;
FILTER15 NEVER;FILTER16 NEVER;
QMESSAGE 1
STATus:CONDition?
Function Queries the contents of the condition filter.
Syntax STATus:CONDition?
Example STATUS:CONDITION -> 16
Description Refer to 14.4 for details on the condition filter.
STATus:EESE
Function Sets the extended event register/queries the
current setting.
Syntax STATus:EESE <Register>
STATus:EESE?
<Register>=0 to 65535
Example STATUS:EESE 257
STATUS:EESE? -> :STATUS:EESE 257
Description Refer to Section 14.4 for details on the
extended event register.
STATus:EESR?
Function Queries the contents of the extended event
register, and clears it.
Syntax STATus:EESR?
Example STATUS:EESR? -> 1
Description Refer to Section 14.4 for details on the
extended event register.
STATus:ERRor?
Function Queries the occurred error code and message.
Syntax STATus:ERRor?
Example STATUS:ERROR?
-> 113,“Undefined header“
STATus:FILTer<x>
Function Sets the transit filter/queries the current setting.
Syntax STATus:FILTer<x> {RISE|FALL|BOTH|
NEVer}
STATus:FILTer<x>?
<x>=1 to 16
Example STATUS:FILTER2 RISE
STATUS:FILTER2?
-> :STATUS:FILTER2 RISE
Description Refer to 14.4 for details on the condition filter.
14.3.13 STATus Group
The commands in the STATus group are used to make settings relating to, and inquiries about the communication
status. There is no corresponding operation using the front panel. Refer to section 14.4 for status reports.
:STATus : CONDition ?
EESE <Space> <Register>
?
EESR ?
ERRor ?
FILTe
r <x> <Space> RISE
FALL
BO
TH
NEV
er
?
QMESsage <Space> OFF
ON
<NRf>
?
SPOLl ?
?
;
14.3 Commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT210 and is the answer not in the manual?

YOKOGAWA WT210 Specifications

General IconGeneral
BrandYOKOGAWA
ModelWT210
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals