16-11
IM 760401-01E
Specifications
16
16.4 External Sensor Input (/EX1 and /EX2 options)
Item Specifications
Allows input of voltage output type current sensor signal. For detailed input specifications, see section 16.1, “Input.”
Measurement range of the /EX1 option:
When the crest factor is set to 3: 2.5 V, 5 V, and 10 V
When the crest factor is set to 6: 1.25 V, 2.5 V, and 5 V
Measurement range of the /EX2 option:
When the crest factor is set to 3: 50 mV, 100 mV, and 200 mV
When the crest factor is set to 6: 25 mV, 50 mV, and 100 mV
16.5 D/A Output (/DA4, /DA12, and /CMP Options)
Item Specifications
Output voltage ±5 V FS (approx. ±7.5 V maximum) against each rated value.
For the relationship between the output items and output voltage (D/A output voltage), see page 9-
8.
Output current ±1 mA
Number of output channels 4 outputs for products with the /DA4 or /CMP option; 12 outputs for products with the /DA12 option.
Output items Set for each channel.
Accuracy ±(accuracy of each measurement item + 0.2% of FS)
Update rate Same as the display update rate.
Temperature coefficient ±0.05%/°C of FS
16.6 Comparator Output (/CMP Option)
Item Specifications
Comparator mode Select single mode or dual mode.
Output type Relay contact output (contact structure: 1C contact)
Number of output relays 4 outputs
Type of limit Set for each relay.
Contact capacity 24 V/0.5 A (rating)
16.7 Remote Control Input/Output Signal (/DA4, /
DA12, and /CMP Options)
Item Specifications
Remote control input signal EXT HOLD, EXT TRIG, EXT START, EXT STOP, EXT RESET
Remote control output signal INTEG BUSY
I/O level TTL
I/O logic format Negative logic, falling edge