7-C-8 Chapter 7: Gas Analyzer Configuration 42/24-10 EN Rev. 9
Replacing a System Module
Removing and
Reinstalling an
Existing System
Module
As a rule, when an existing system module is removed and reinstalled (e.g. after a
repair) it does not require any setup.
When the system module is reconnected to the system bus it is automatically
recognized and its configuration is automatically stored. For automatic recognition
to take place the gas analyzer must be in measurement mode.
CAUTION!
When an existing system module is replaced with another system module the
function “DELETE” should not be used to delete the old system module. This
function would irretrievably delete the parameter settings and function block
configuration of the old system module!
In order to preserve the parameter settings and function block configuration
of the old system module when replacing a system module, the function
“CHANGE” must be used.
The type and configuration of the new system module must match the charac-
teristics of the old system module.
When an existing I/O module has been replaced with an I/O module of the same
type, the new I/O module is automatically recognized by the gas analyzer and its
configuration is not necessary.
Step Action
1 Select the Set Up System Modules menu item.
The list of the system modules existing in the system is displayed.
2 Select the system module (analyzer module or I/O device) to be re-
placed and reconfigured. In the list this system module will be shown
with an Unknown or Error status.
Press the CHANGE softkey. 3
Do not press the DELETE softkey! This will irretrievably
remove the system module’s parameter settings and function
block configuration!
4 Enter the 14-digit serial number for the new system module.
5 In the list the new system module will now be shown with a
Replaced status.
Replacing an Existing
System Module
(Analyzer Module or
I/O Device) with
Another System
Module
6 Save the configuration change by pressing ENTER or discard it by
pressing Back.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB Advance Optima magnos206 and is the answer not in the manual?

ABB Advance Optima magnos206 Specifications

General IconGeneral
BrandABB
ModelAdvance Optima magnos206
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals