42/24-10 EN Rev. 9 Chapter 10: Status Messages, Troubleshooting 10-23
Gas Module Problems
Flow Problem
Condensation in the
flow meter
Disconnect the gas analyzer from the gas preparation system.
Dry the flow meter by heating it and blowing it with compressed air.
Check the operation of the upstream sample gas cooler.
Inadequate gas
supply
Connect the flow meter, ball chamber or pressure gauge directly to the gas
supply pump and measure the pressure or vacuum.
Check the pump and change the membrane as necessary.
Check and, if necessary, replace the disposable filter (see page 9-24 for
instructions).
Check and, if necessary, replace the solenoid(s).