42/24-10 EN Rev. 9 Appendix 2: Analyzer Module Operating Specifications A-2-7
Oxygen Sensor Operating Specifications
Linearity deviation Linear in the range > 1 Vol.-% O
2
Repeatability 0.5 % of span
Zero drift Stable over long-term due to absolute zero point
Sensitivity drift 1 % of measurement range scale per week
Output fluctuation (2 σ) 0.2 % of the measurement range at electronic T90 time (static/dynamic)
= 5/ 0 sec.
Detection limit (4 σ) 0.4 % of the measurement range at electronic T90 time (static/dynamic)
= 5/ 0 sec.
Flow influence Flow rate in 20 to 100 l/h range: 2 % of measurement range scale
Temperature
influence
Ambient temperature in permissible range: 0.2 Vol.-% O
2
per 10 °C
Air pressure influence At zero-point: No effect
On sensitivity without pressure correction:
1 % of measured value per 1 % air
pressure change
On sensitivity with pressure correction:
0.2 % of measured value per 1 % air
pressure change
Pressure correction is only possible if the oxygen sensor is connected to an
analyzer module with an integral pressure sensor.
Pressure sensor working range: p
abs
= 600 to 1250 hPa
T
90
time T
90
30 seconds, depending on sample gas flow and system layout

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB Advance Optima magnos206 and is the answer not in the manual?

ABB Advance Optima magnos206 Specifications

General IconGeneral
BrandABB
ModelAdvance Optima magnos206
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals