8-B-2 Chapter 8: Gas Analyzer Calibration 42/24-10 EN Rev. 9
Calibration Data for Manual Calibration
Menu Path MENU → Configure → Calibration data →
Manual calibration → ...
Test Gas
Concentration
The zero and span test gas concentrations to be used as set points for manual
calibration need to be set for the selected sample component and measurement
range.
Calibration Method The method for manual calibration needs to be set (see also “Calibration
Methods” section, page 8-A-10).
For ... ... select:
Common calibration • The sample component and
• The measurement ranges for start- and end-point
calibration for the selected component.
Substitute gas calibration • The (substitute gas) components for start- and
end-point calibration and
• The measurement range for the selected
component.
The components and measurement ranges settings apply to manual, automatic
and externally controlled calibration.