EasyManuals Logo

ABB Advance Optima magnos206 User Manual

ABB Advance Optima magnos206
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #156 background imageLoading...
Page #156 background image
8-C-2 Chapter 8: Gas Analyzer Calibration 42/24-10 EN Rev. 9
Caldos27: Notes for Calibrating
Sample Components The Caldos27 analyzer module has at least one sample component with one
measurement range and, for measurement ranges Class 1, the standard gas
substitute gas component.
Associated Gas
Influence
The Caldos27 analyzer module measurement technique is based on the differing
thermal conductivity of various gases.
Since this technique is non-selective, the concentration of a sample component
can be accurately measured only in a binary or quasi-binary gas mixture.
If other associated gas components are present in the sample gas their effect on
factory calibration must be considered.
Test Gases Zero Calibration: Test gas or sample-component-free process gas or substitute
gas
Span Calibration: Test gas or process gas having a known sample gas concen-
tration or substitute gas
Test Gas for
Corrected Sample
Components
During calibration cross-sensitivity and/or carrier gas corrections by other meas-
urement components are switched off. Therefore, a corrected measurement com-
ponent should be calibrated only using a test gas consisting of the measurement
component and an inert gas like N
2
.
Calibration with
Substitute Gases
If test gases are not available for calibration, the Caldos27 analyzer module can
be factory-set for calibration with substitute gases (see the “Calibration Methods”
section, page 8-A-10).
Substitute gas calibration of the Caldos27 analyzer module is described using the
example of “CO
2
Measurement in Flue Gas”, on page 8-C-4.
Single-Point
Calibration with
Standard Gas
Single-point calibration of the Caldos27 analyzer module with standard gas is
described on page 8-C-3.
Sequence of
Calibration
Zero point calibration must always precede end point calibration.
Wait Until End of
Warm-Up Phase
The Caldos27 analyzer module should only be calibrated after the warm-up
phase, i.e. approx. 30 minutes after connecting the power supply.
The time required for the warm-up phase depends on the measurement range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB Advance Optima magnos206 and is the answer not in the manual?

ABB Advance Optima magnos206 Specifications

General IconGeneral
BrandABB
ModelAdvance Optima magnos206
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals