EasyManuals Logo

ABB Advance Optima magnos206 User Manual

ABB Advance Optima magnos206
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #176 background imageLoading...
Page #176 background image
9-4 Chapter 9: Inspection and Maintenance 42/24-10 EN Rev. 9
Uras26: Optical Alignment
Definition Optical alignment of the Uras26 analyzer module will minimize asymmetry of the
radiation which falls in through the sample and reference sides of the sample cell.
When should optical
alignment be
performed?
An optical alignment must always be performed
If the offset drift has fallen below the permissible range (50% of the physical
measurement range)
After a component (emitter, sample cell, calibration unit /cell, detector) is
installed in or removed from a beam path.
How should optical
alignment be
performed?
Each beam path in the analyzer module must be optically aligned separately. If
there are two detectors in a beam path, the optical alignment should be
performed in the rearmost detector (as seen from the emitter). During an optical
alignment, beam path intensity is varied by means of mechanical apertures and, if
necessary, by turning the emitter case. For this, the system housing must be
opened (see Figure 9-4, page 9-5).
Detector
Arrangement
Figure 9-3 shows the detector layout. Detector numbers correspond to the
numbers assigned to the series of measurement components as shown on the
analyzer module identification plate.
Figure 9-3
Uras26 Analyzer
Module Detector
Numbering
11
2
1 1
2
1
3
1
42
Test Gas During optical alignment the zero gas supply must be turned on.
Emitter Wrench An “emitter wrench” is required to rotate the emitter case (see Figure 9-4, page
9-5). It is affixed to the analyzer module.
CAUTION!
Current-bearing components can be exposed when the covers or parts are
removed, even if this can be done without tools. Current can be present at
some connection points. All work on a gas analyzer that is open and con-
nected to power should only be performed by trained personnel who are
familiar with the risks involved.
Menu Path MENU
Maintenance/Test Analyzer spec. adjustm.
Optical adjustm.
Continued on next page

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB Advance Optima magnos206 and is the answer not in the manual?

ABB Advance Optima magnos206 Specifications

General IconGeneral
BrandABB
ModelAdvance Optima magnos206
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals