Defining the System Setup: Define the ECG Setup
MAC 5000 System — 2000657-057, Revision B 14-23
Extra Leads (15 Lead ECG Option Only)
Vector Loops (Vector Loops Option Only)
Analog Outputs
Select this function to define the system’s output signals
when connecting additional equipment to the system.
Item Description
Lead Set Choose the three additional leads used. You can:
n
select one of the pre-defined lead sets, or
n
select Custom 3 to define the electrode positions of
A1, A2, and A3.
Item Description
Number of
copies
Type a value between 0 and 10 for the number of
copies you want to print for this report format.
Main loop gain Select a default setting.
Lead Z display Select a default setting.
Sagittal plane Select a default setting.
Item Description
Fast Analog
Output
Select Not Used, I, II, or V1-V6.
TTL Output Select Not Used or QRS Detect to define TTL Output.
Polarity Select Positive or Negative to define TTL Output
polarity.
Width Type a value between 0 and 48 to define TTL Output
signal width in milliseconds.
Delay Type a value between 0 and 100 to set a delay in
milliseconds for the TTL Output QRS detector signal.
QRS Beep Select On to hear a beep for each QRS complex.
System Setup
ECG
ECG
Analog Outputs
168A, 057A
To Purchase, Visit Avobus.com or call 1-800-674-3655

Table of Contents

Other manuals for GE MAC 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5000 and is the answer not in the manual?

GE MAC 5000 Specifications

General IconGeneral
BrandGE
ModelMAC 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals