Appendix F – Master’s Step Data: ST-T Change
MAC 5000 System — 2000657-057, Revision B F-5
ST-T Change
The existence of any ST-T change is assessed by classifying
ST-T into three assessment levels: positive, borderline, and
negative. The following criteria is used:
Positive
n
ST depression 0.1mV (2 or more leads).
n
ST elevation 0.2mV (2 or more leads).
n
T wave change 1.0mV (2 or more leads).
u
No. 1, 2, or 3 is satisfied.
Borderline
n
ST depression 0.05mV (any leads).
n
ST elevation 0.1mV (any leads).
n
T wave change 0.5mV (any leads).
u
No. 1, 2, or 3 is satisfied.
Negative
n
Positive and borderline criteria are NOT satisfied.
Calculation
ST depression = (rest ST - post J) - (post exercise ST -
post J)
ST elevation = (post exercise ST - post J) - (rest ST - post J)
T wave change = absolute value of (rest T wave amplitude -
post-exercise T wave amplitude)
(ST - post J: amplitude at the post J point)
When the assessment is positive or borderline, the lead
with the largest change prints.

Table of Contents

Other manuals for GE MAC 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5000 and is the answer not in the manual?

GE MAC 5000 Specifications

General IconGeneral
BrandGE
ModelMAC 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals