2-18 Output Connections Model 6220/6221 Reference Manual
Return to Section 2 topics
Figure 2-14
Cable Guard connections – triax inner shield connected to Cable Guard
Using shielding and guarding together
Figure 2-15 shows connections for a test system that uses a noise shield, a safety
shield, and guarding.
Figure 2-15
Connections for noise shield, safety shield, and guarding
High
Low
DUT
Output
622x
Guard
Plate
Cable Guard
(black*)
Triax Cable
Output Low connected
to earth ground
(red*)
(green*)
* Boot color for alligator clip if
using supplied triax cable.
(Earth Ground)
High
Low
DUT
Output
Metal Noise Shield
Safety
Earth
Ground
Metal
Safety
Shield
Chassis
Screw
622x
Guard
Plate
Cable Guard
(black*)
Triax Cable
Triax inner shield connected to Cable Guard.
Output Low connected to earth ground.
(red*)
(green*)
* Boot color for alligator clip if
using supplied triax cable.
(Earth Ground)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals