Safety Precautions
1.4.6
Precautions Following Requirements of Laws and Regulations
CAUTION
• Caution as to laser control
To prevent danger, make sure that only service engineers who received proper formal training will remove the
exterior cover or touch the inside of the equipment.
DRYPRO 873 Laser Unit Specications
Item Specifications
Wavelength
784nm
Maximum output
280mW
(1) EMC Statement
This device has been tested and found to comply with the limits for medical devices in IEC 60601-
1-2: 2007 or IEC 60601-1-2: 2014. These limits are designed to provide reasonable protection
against harmful interference in a typical medical installation. DRYPRO 873 generates, uses and
can radiate radio frequency energy and, if not installed and used in accordance with the instruc
-
tions, may cause harmful interference to other devices in the vicinity. However, there is no guaran-
tee that interference will not occur in a particular installation. If DRYPRO 873 does cause harmful
interference to other devices, which can be determined by turning the equipment o and on, the
user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit dierent from that to which the other device(s)
are connected.
- Consult the manufacturer or eld service technician for help.
•• WARNING
• Do not use mobile RF communication devices such as mobile phones or pocket pagers in the vicinity of this device.
Use of such RF devices near this device can cause errors in operation due to electromagnetic wave interference.
When using a mobile RF device, ensure that the distance between the RF device in use and this device is at least
30 cm (12 inchies).
• The use of accessories, transducers and cables other than those sold by Konica Minolta, Inc. as internal compo-
nents, may result in increased emissions or decreased electromagnetic immunity of this device, possibly causing
malfunction.
• Do not use this device adjacent to or stacked with other devices. If adjacent or stacked use is necessary, conrm
normal operation in the conguration in which this device will be used.
•• Supplementary information regarding IEC 60601-1-2: 2007 and IEC 60601-1-2: 2014
(1) Take precautions against this device especially regarding EMC. Install and put into service accord-
ing to the electromagnetic compatibility (EMC) information provided in the manual (Table 1 - Table
5).
(2) This device is suitable for use in medical institutions such as hospitals and clinics. However, this
device is not suitable for use in environments with high degree of electromagnetic wave interfer-
ence (such as near an electric operation device in use for output, or near a high frequency treat-
ment device).
(3) Cable List-
AC Power Cable (3m)