4.6 How to Handle Film Flaws
How to Handle Film Flaws That Consistently Occur in the Same Position
If aws consistently occur in the same position on printed lm, it may be that foreign matter has been
deposited on the feed guides.
In such a case, clean the feed guides with a soft lint-free cloth dampened with ethanol. The feed guides
in the following 2 locations must be cleaned.
Inside the top right cover
Inside the lower right cover (when an expansion tray or mammography media expansion tray is at
-
tached)
Wiping the feed guides too harshly may damage them, which in
turn would cause more lm aws. Therefore, wipe them carefully.
If lm aws a󰀨ecting the interpretation of radiographs persist even after
cleaning the feed guides, contact Konica Minolta technical representatives.
Cleaning the feed guides inside the top right cover
Open the top right cover, and wipe the guide plate and roller with an ethanol-dampened cloth.
Cleaning the feed guides inside the lower right cover (when an expansion
tray or mammography media expansion tray is attached)
Open the lower right cover, and wipe the guide plate and roller with an ethanol-dampened cloth.

Table of Contents

Other manuals for Konica Minolta DRYPRO 873

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta DRYPRO 873 and is the answer not in the manual?

Konica Minolta DRYPRO 873 Specifications

General IconGeneral
BrandKonica Minolta
ModelDRYPRO 873
CategoryMedical Equipment
LanguageEnglish

Related product manuals