EasyManuals Logo
Home>Mindray>Medical Equipment>BeneHeart D2

Mindray BeneHeart D2 User Manual

Mindray BeneHeart D2
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
Defibrillator/Monitor Operator’s Manual 14 - 1
14 Monitoring CO
2
14.1 CO
2
Introduction
CO
2
monitoring is a continuous, non-invasive technique for determining the concentration of CO
2
in the
patient’s airway by measuring the absorption of infrared (IR) light of specific wavelengths. The CO
2
has its own
absorption characteristic and the amount of light passing the gas probe depends on the concentration of the
measured CO
2
. When a specific band of IR light is passed through respiratory gas samples, some of IR light will
be absorbed by the CO
2
molecules. The amount of IR light transmitted after it has been passed through the
respiratory gas sample is measured with a photodetector. From the amount of IR light measured, the
concentration of CO
2
is calculated.
CO
2
monitoring is intended for adult, pediatric and neonatal patients.
14.2 CO
2
Safety Information
WARNING
• Remove the airway sampling line from the patient’s airway while nebulized medications are being
delivered.
• Leakage in the breathing or sampling system may cause the displayed EtCO
2
values to be
significantly low. Always make sure that all components are securely connected.
• EtCO
2
values measured from the CO
2
module may differ from those of from the blood gas analysis.
• Route all tubing away from the patient’s throat to avoid strangulation.
• Inspect the airway adapter for a tight connection and proper operation before attaching it to the
patient.
• Squeezing or bending the sampling line during the CO
2
measurement may cause inaccurate CO
2
reading or no reading.
14.3 CO
2
Measurement Limitations
The following factors may influence the accuracy of measurement:
â–  Leakage or internal venting of sampled gas
â–  Mechanical shock
â–  Cyclic pressure higher than 10 kPa (100 cmH
2
O) or abnormal change to airway pressure
â–  Other sources of interference, if any.
Measurement accuracy of the CO
2
m
odule may be affected by the breath rate and inspiration/expiration (I/E)
ratio. For details, refer to A.5.7CO
2
Specifications.

Table of Contents

Other manuals for Mindray BeneHeart D2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D2 and is the answer not in the manual?

Mindray BeneHeart D2 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D2
CategoryMedical Equipment
LanguageEnglish

Related product manuals