Defibrillator/Monitor Operator’s Manual 21 - 9
21.6.9 PR Setup Menu
21.6.10 NIBP Setup Menu
Menu Item Options/Range Default Description
PR Alarm On, Off Off /
PR Alm Lev High, Med, Low Med /
PR High Adu (Low+2) to 300 120 Step: 1 bpm
PR ranges are different for
different modules.
Ped 160
Neo 200
PR Low Adu 15 to (High-2) 50
Ped 75
Neo 100
QRS Volume 0 to 10 2 This setting is linked with the
[QRS Volume] setting in the
[ECG Setup] menu
Menu Item Options/Range Default Description
Interval Manual, 1 min, 2 min, 2.5
min, 3 min, 5 min, 10 min, 15
min, 20 min, 30 min, 1 h, 1.5
h, 2 h, 3 h, 4 h, 8 h
Manual Sets the interval between NIBP
measurements.
Press. Unit mmHg, kPa mmHg Sets the NIBP measurement
unit.
Alarm limits are refreshed in real
time when pressure unit is
changed.
Alarm NIBP-Sys On, Off On /
NIBP-Dia
NIBP-Mean
Alm Lev High, Med Med /
Sys High Adu (Low+5) to 290 160 Step: 1 mmHg
Ped (Low+5) to 240 120
Neo (Low+5) to 140 90
Sys Low Adu 25 to (High-5) 90
Ped 70
Neo 40
Mean High Adu (Low+5) to 260 110
Ped (Low+5) to 215 90
Neo (Low+5) to 125 70
Mean Low Adu 15 to (High-5) 60
Ped 50
Neo 25

Table of Contents

Other manuals for Mindray BeneHeart D2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D2 and is the answer not in the manual?

Mindray BeneHeart D2 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D2
CategoryMedical Equipment
LanguageEnglish

Related product manuals