A - 14 Defibrillator/Monitor Operator’s Manual
A.7 Recorder Specifications
A.8 Alarm Specifications
A.9 Data Storage
Method High-resolution thermal dot array
Number of waveforms 3 at maximum
Paper speed 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s. Error no more than ± 5%
Paper width 50 mm
Grid lines The operator can choose to print grid lines or not
Auto record Charge events, shock events, marked events, auto test report, parameter
alarms, ARR alarms, if configured on
Alarm Levels High, medium, low level alarms, complying with IEC60601-1-8
Alarm Categories Physiological alarms, technical alarms; Latched alarms and unlatched
alarms.
Alarm lamp Independent alarm LED
Parameter alarm setting Alarm properties of all available parameters can be set simultaneously in the
Para. Alarm menu
Auto alarm limits Parameter alarm limits can be automatically adjusted according to currently
measured vital signs
Internal storage 1G Bytes
Marking events 16 types of events, user customized
Events At least 1000 events for each patient.
Waveforms At least 24 hours of consecutive ECG waveform
Voice recording
(available in the AED mode)
At least 180 minutes in total, more than 60 minutes for each patient
Data export Data can be export to a PC through a USB flash memory
Tabular trends Up to 72 hours for all measured parameters with the resolution no less than
1 minute.
Patient archives Up to 100