Defibrillator/Monitor Operator’s Manual A - 13
Note: The specifications above are based on a new battery, and at 20ºC±5 °C of ambient temperature.
Battery LI24I001A charge
time
Charged by the equipment
connected to the AC power
Less than 2 hours to 90% and less than 3 hours to 100%
with equipment power off;
Less than 3.5 hours to 90% and less than 4.5 hours to
100% with equipment power on.
Charged by the charger station Less than 2.5 hours to 90% and less than 3 hours to
100%.
Operating
mode
One new fully
charged
battery
Testing condition
Battery LI24I005A run time Monitoring ≥6 h The equipment is configured with a 5-lead ECG, Resp,
SpO
2
and CO
2
, NIBP measurements set at an interval of
15 minutes. WI-Fi is disabled. The screen brightness is
set to the factory default without recording.
Defibrillation ≥200
discharges
360J discharge, the equipment is configured with a 5-
lead ECG, Resp, SpO
2
, NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to the factory default
without recording.
Defibrillation ≥300
discharges
200J discharge, the equipment is configured with a 5-
lead ECG, Resp, SpO
2
, NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to the factory default
without recording.
Pacing ≥4.5 h 50  load impedance, pacing rate 80ppm, pacing
output 60mA. The equipment is configured with a 5-
lead ECG, Resp, SpO
2
, NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to the factory default
without recording.
Operating
mode
One new fully
charged
battery
Testing condition
Battery LI24I001A run time Monitoring ≥3 h The equipment is configured with a 5-lead ECG, Resp,
SpO
2
and CO
2
, NIBP measurements set at an interval of
15 minutes.
WI-Fi is disabled. The screen brightness is
set to the factory default without recording.
Defibrillation ≥100
discharges
360J discharge at intervals of one minute, without
recording
Pacing ≥2 h 50  load impedance, pacing rate 80ppm, pacing
output 60mA, without recording
Battery fuel gauge 5 LEDs indicating the current battery charge level
Shutdown delay At least 20 minutes of monitoring and six 360J discharges (after the low battery alarm
occurs)