24 - 6 Defibrillator/Monitor Operator’s Manual
Energy Disarming
1. Disconnect the equipment from the AC mains, and connect the equipment with fully charged battery.
2. Switch the Mode Select knob to Manual Defib.
3. Connect the electrode pads/external paddles to the equipment, and place the electrode pads/external
paddles on the defibrillator/pacer analyzer.
4. Set the analyzer to Energy Measurement mode, and check that the energy value is 0 or blank.
5. Select the energy level to 360J.
6. Charge the equipment, and check that the charge tone is issued during charging.
7. Press the [Disarm] soft key to discharge the energy internally.
8. Check that a prompt “Charge Removed appears on the screen and the charge done tone stops, the value
measured by the analyzer is 0J or blank.
9. Press the Main Menu button on the front panel, and then select [Others >>] [Configuration >>]
enter the required password.
10. Select [Manual Defib Setup] and set [Time to Auto Disarm] to [60s].
11. Set the analyzer to Energy Measurement mode, and check that the energy value is 0 or blank.
12. Select the energy level to 360J.
13. Charge the equipment. After charging is completed, wait 60 seconds to check that the prompt “Charge
Removed appears on the equipment and the energy measured by the analyzer is 0J or blank.
Synchronized Cardioversion
1. Connect the electrode pads/external paddles to the equipment, and place the electrode pads/external
paddles on the defibrillator/pacer analyzer.
2. Connect the ECG cable to the equipment, and place the ECG electrodes on the defibrillator/pacer analyzer.
3. Set the analyzer to Time Measurement Mode and output normal sinus rhythms, e.g. amplitude value 1mV
and HR 60bpm.
4. Press the Main Menu button on the front panel, and then select [Others >>] [Configuration >>]
enter the required password.
5. Select [Manual Defib Setup] and set [Sync After Shock] to [On].
6. Select the energy level to 10J.
7. Press the [Enter Sync] soft key to start synchronized cardioversion. If Remote Sync is switched on, press the
[Enter Sync] soft key and select [Local] to start synchronized cardioversion.
8. Select Pads or Paddles as the ECG source, and charge the equipment.
9. Press the Shock button to deliver a shock. Check the following items:
The synchronized discharge succeeds and the delivered energy measured by the analyzer is 10J±2J.
The delay time of synchronized cardioversion measured by the analyzer is less than 60ms.
The synchronized discharge mark appears on the R wave.
The prompt messages are correct during the test.
10. Select lead II as ECG source, and charge the equipment. Repeat step 9.
24.4.4 Pacing Test
Test tools: defibrillator/pacer analyzer
1. Disconnect the equipment from the AC mains, and connect the equipment with fully charged battery.
2. Switch the Mode Select knob to Pacer.
3. Select the Fixed mode.
4. Connect the pads cable to the equipment, and place the pads on the defibrillator/pacer analyzer.
5. Set the analyzer to Pacing Measurement mode. Use test load of 50Ω.
6. Set [Pacer Rate] to [70ppm] and [Pacer Output] to [30mA].

Table of Contents

Other manuals for Mindray BeneHeart D2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D2 and is the answer not in the manual?

Mindray BeneHeart D2 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D2
CategoryMedical Equipment
LanguageEnglish

Related product manuals