21 - 8 Defibrillator/Monitor Operator’s Manual
21.6.7 Resp Setup Menu
21.6.8 SpO
2
Setup Menu
Menu Item Options/Range Default Description
RR Alarm On, Off Off /
RR Alm Lev High, Med, Low Med /
Sweep 6.25 mm/s, 12.5 mm/s, 25
mm/s
6.25 mm/s Sets the Resp waveform speed.
RR High Adu (Low+2) to 100 30 Step: 1 rpm
Ped 30
Neo (Low+2) to 150 100
RR Low Adu 6 to (High-2) 8
Ped 8
Neo 30
Apnea Time 10 s, 15s, 20s, 25s, 30s, 35s,
40s
20 s /
Menu Item Options/Range Default Description
SpO2 Alarm On, Off On /
SpO2 Alm Lev High, Med Med /
Sweep 12.5 mm/s, 25 mm/s 25 mm/s Sets the SpO
2
waveform speed.
SpO2 High Adu (Low+1) to 100 100 Step: 1%
Ped 100
Neo 95
SpO2 Low Adu Desat Limit to
(High-1)
90
Ped 90
Neo 90
Desat Adu 50 to (High-1) 80
Ped 80
Neo 80
Pitch Tone Mode 1, Mode 2 Mode 1 Sets the SpO
2
tone mode
.
Sat-Seconds 0s, 10s, 25s, 50s, 100s 0s For Nellcor SpO
2
module only.
Sensitivity Mindray SpO
2
High, Med, Low Med Different options are available
to match the SpO
2
module
used.
Masimo SpO
2
High, Normal, APOD Normal
Average Time 2-4s, 4-6s, 8s, 10s, 12s, 14s,
16s
8s For Masimo SpO
2
module only.
Fast SAT On, Off Off
Display SIQ On, Off Off
Alarm Delay 0s to 15s 6s

Table of Contents

Other manuals for Mindray BeneHeart D2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D2 and is the answer not in the manual?

Mindray BeneHeart D2 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D2
CategoryMedical Equipment
LanguageEnglish

Related product manuals