EasyManuals Logo
Home>Mindray>Medical Equipment>BeneHeart D2

Mindray BeneHeart D2 User Manual

Mindray BeneHeart D2
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
Defibrillator/Monitor Operator’s Manual 21 - 1
21 Configuration Management
21.1 Configuration Management Introduction
The configuration management enables you to customize you equipment to best meet your needs. With this
function, you can:
â–  View system configurations
â–  Change system configurations
â–  Record system configurations
â–  Export configurations
â–  Import configurations
â–  Restore the factory default configurations
After the system configurations have been changed, the equipment restarts and new configuration settings take
effect immediately.
WARNING
• Entering the Configuration Main menu is password protected. Patient monitoring and therapy
automatically ends when you access the configuration management.
• The configuration management tasks must be performed by clinical professionals.
• Never connect the equipment with the patient when accessing the configuration management.
21.2 Modifying Configuration Management Password
The required password for entering the Configuration Main menu is set to 315666.
You can also modify the configuration password.
1. Press the Main Menu button on the front panel, and then select [Others >>] → [Configuration >>] →
enter the required password.
2. Select [Others >>] → [Modify Password].
3. Respectively input the old password and new password. The maximum length of a new password can be
set to 32 characters.
4. Select [Confirm].
21.3 Viewing Configurations
Password is not required when you view configurations.
1. Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
1. Enter the Manual Defib mode, the Monitor mode, or the Pacer mode.
2. Press the Main Menu button on the front panel, and then select [Others >>] → [Configuration >>].
â—† Selecting [View Config] can view the configurations.
◆ Selecting [View Config] → [General Setup] can change the system time.
◆ Selecting [View Config] → [Record] can record all the configurations.
â—† Selecting [Cancel] can close the dialog box and return to the normal operating mode.

Table of Contents

Other manuals for Mindray BeneHeart D2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D2 and is the answer not in the manual?

Mindray BeneHeart D2 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D2
CategoryMedical Equipment
LanguageEnglish

Related product manuals