C - 4 Defibrillator/Monitor Operator’s Manual
C.2 Radio Regulatory Compliance
The device comply with the essential requirements and other relevant provisions of Directive 2014/53/EU.
• Keep a distance of at least 20 cm away from the equipment when wireless function is in use.
Recommended Separation Distances between Portable and Mobile RF, Communications Equipment and This
Equipment
The equipment is intended for use in an electromagnetic environment in which radiated RF disturbance are controlled.
The customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the equipment as
recommended below, according to the maximum output power of the communication equipment.
Rated Maximum Output
power of Transmitter Watts
(W)
Separation Distance According to Frequency of Transmitter (m)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.7 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.20 1.20 2.30
10 3.80 3.80 7.30
100 12.00 12.00 23.00
For transmitters at a maximum output power not listed above, the recommended separation distanced in meters (m) can
be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
P
V
d






=
1
5.3
P
E
d






=
1
5.3