SpO
2
Monitoring
12-19
îš„ PR SOUND
Pulse volume
Range: 0–10. 0 indicates the volume is closed and 10
indicates the maximum volume.
îš„ AV G T I M E
Average time
Determines the average SpO
2
calculation time.
Options: 2-4S, 4-6S, 8S, 10S, 12S, 14S and 16S.
îš„ SENSITIVITY
MODE
Options: NORMAL and HIGH.
îš„ NIBP SIMUL
ON: When measuring NIBP and SpO
2
at the same side, the
SpO
2
alarm status remains unchanged until the NIBP
measurement is finished;
OFF: The function of NIBP SIMUL is disabled.
îš„ DEFAULT
You can select this option to access the SPO2 DEFAULT
CONFIG menu, in which you may select FACTORY
DEFAULT CONFIG or USER DEFAULT CONFIG. After
finishing your selection and exiting the menu, the system
pops up a dialog box asking for your confirmation.
12.3.6 Sensors and Accessories
If your monitor is equipped with a Masimo SpO
2
module, use only Masimo oximetry
sensors for SpO
2
measurements. Other sensors may cause improper pulse oximeter
performance. Before use, carefully read the directions for the LNOP sensor. Tissue
damage can be caused by incorrect application or use of a sensor. An example is
wrapping the sensor too tightly. Inspect the sensor site as specified in the directions to
ensure skin integrity, correct positioning, and adhesion of the sensor.
CAUTION
z Do not use damaged sensors. Do not use a sensor with exposed optical
components. Do not immerse the sensor in water, solvents, or cleaning
solutions (the sensors and connectors are not waterproof). Do not
sterilize by irradiation, steam, or ethylene oxide.
z Do not use damaged patient cables. Do not immerse the patient cables
in water, solvents, or cleaning solutions (the sensors and connectors are
not waterproof). Do not sterilize by irradiation, steam, or ethylene oxide.