Appendices
20-13
A.11 NIBP Specifications
Measurement technique Auto oscillation
Displayed parameters Systolic pressure, diastolic pressure and mean pressure
Mode of operation Manual, auto and continuous
Measurement interval in
auto mode
1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes
Measurement time in
continuous mode
5 minutes
mmHg Adult Pediatric Neonate
Systolic
pressure
40 to 270 40 to 200 40 to 135
Diastolic
pressure
10 to 210 10 to 150 10 to 100
Measurement range in
normal mode
Mean pressure 20 to 230 20 to 165 20 to 110
Measurement precision
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg
Resolution 1mmHg
Over-pressure protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg