17.3.2 Preparations for CO
2
Measurement
1. Plug the sampling line into its receptacle before the measurement.
2. Open the CO2 SETUP menu and set WORK MODE to MEASURE.
3. The “CO2 START UP” prompt information is displayed on the screen until the
startup is finished.
Figure 17-7 CO
2
Module Airway Connections
When the monitor is powered on for the first time, the CO
2
module enters the
STANDBY mode by default. To activat the CO
2
module, you must change the work
mode to MEASURE. The work mode of the CO
2
module keeps unchanged when the
monitor is restarted. For instance, if the monitor is power off when the CO
2
module
is working in the MEASURE mode, the CO
2
module will automatically enter the
MEASURE mode after the monitor is restarted. For details about WORK MODE,
please refer to the section 17.3.3 CO2 .
WARNING
z Do not use the accessory if the packaging or the internal accessory is
damaged. Return it to the manufacturer.
z The sidestream sampling line is disposable. It should not be disinfected
for reuse or cross-used by different patients.