IBP Monitoring
15-14
15.5 Maintenance and Cleaning
WARNING
z Before cleaning the transducer, make sure the transducer is
disconnected from the monitor, or the monitor is powered off and
disconnected from AC power.
Cleaning of IBP Transducer
After the IBP monitoring operation is completed, remove the tubing and the dome
from the transducer and wipe the transducer diaphragm with water. To clean the
transducer and the cable, wipe them using soap or the detergents listed below:
îš„ Cetylcide
îš„ Wavicide-01
îš„ Wescodyne
îš„ Cidex
îš„ Lysol
îš„ Vesp he ne
Do not immerse the connector in any liquid. After cleaning, dry the transducer
thoroughly before storing. Slight discoloration or temporary increase of surface
stickiness of the cable should not be considered abnormal. If it is necessary to
remove the adhesive tape residue from the transducer cable, the double seal tape
remover is recommended; use the remover with special caution to minimize the
damage to the cable. Acetone, Alcohol, Ammonia and Chloroform, or other strong
solvents are not recommended because they are harmful to the vinyl cabling if used
for a long time.
NOTE
z The disposable transducers or domes must not be reused.
z To avoid contamination, the disposable transducers or domes must be
reclaimed or disposed of properly.