CO
2
Monitoring
17-6
17.2.3 CO
2
Setup Menu
Selecting the CO
2
label in the parameter window opens the following menu.
Figure 17-3 CO
2
Setup Menu
You can perform the following settings in this menu.
îš„ ALM
Alarm on/off status
ON: When a CO
2
alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a CO
2
alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the
right of the CO
2
label.
îš„ ALM LEV
Alarm level
Options: HIGH, MED and LOW.
îš„ ALM REC
Alarm recording
ON: When a CO
2
alarm occurs, the monitor enables the
recording;
OFF: When a CO
2
alarm occurs, the monitor does not
enable the recording.
îš„ CO2 ALM HI
Upper alarm limit of EtCO
2
îš„ CO2 ALM LO
Lower alarm limit of EtCO
2
îš„ INS ALM HI
Upper alarm limit of InsCO
2