17.4.2 Preparations for CO
2
Measurement
1. Plug the water trap into its receptacle before the measurement.
2. Power on the monitor and open the CO2 SETUP menu. In the menu, set the
WORK MODE to MEASURE.
3. The "CO2 WARM UP" technical information is displayed on the screen until
the sensor reaches its work temperature.
Figure 17-12 CO
2
Module Airway Connections
When the monitor is powered on for the first time, the CO
2
module enters the
STANDBY mode by default. To activat the CO
2
module, you must change the work
mode to MEASURE. The work mode of the CO
2
module keeps unchanged when the
monitor is restarted. For instance, if the monitor is power off when the CO
2
module
is working in the MEASURE mode, the CO
2
module will automatically enter the
MEASURE mode after the monitor is restarted. For details about WORK MODE,
please refer to the section Other Setup of 17.4.3 CO2 Setup Menu.
Mainstream
Senso
r
Airway
Adapter

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray PM-9000 and is the answer not in the manual?

Mindray PM-9000 Specifications

General IconGeneral
BrandMindray
ModelPM-9000
CategoryMedical Equipment
LanguageEnglish

Related product manuals