5.6 Standby Mode
During patient transport or temporary departure of a patient, the monitor can be set
to STANDBY mode. In this mode, the monitor suspends the monitoring and
measurement on the patient and shields all alarm indications. Besides the WORK
MODE of CO
2
and AG module, which will also be changed to STANDBY, the
previous menu settings and patient information keep unchanged.
Entering the STANDBY mode
1. Disconnect all leads and sensors between the patient and the monitor.
2. Select the STANDBY label at the lower right corner of the main screen. A
dialog box pops up, and you can choose to enter the standby mode or not.
3. Select YES, and the monitor will enter the standby mode as shown below.
4. Select NO, and the monitor will return to the previous screen.
Figure 5-7 Standby Mode
Exiting the STANDBY mode
Press any key other than the power switch on the control panel or turn the control
knob in the STANDBY mode, and a dialog box will pop up. Select YES in the
dialog box, and you will exit the STANDBY mode and return to the previous screen.