Chapter 3 Installation and Connection
25
VIDEO SYSTEM OTV-SI
Chapter 3 Installation and Connection
• Review this chapter thoroughly before each use. If the
equipment is not properly prepared before each use,
equipment damage, patient and operator injury and/or fire
can occur.
• When non-medical electrical ancillary equipment is being
used, connect the power cords via an isolation transformer
prior to connecting the signal cables. Failure to do this can
cause an electric shock, burns and/or fire.
• Turn OFF all system components before connecting them.
Otherwise, equipment damage can result.
• Use appropriate cables. Otherwise, equipment damage or
malfunction can result.
• When plugging connectors into terminals, hold the connector
plug horizontal to the terminal and push it in completely
without applying excessive force. If the connector is held at
an angle, or if excessive force is used to connect it, the
connector pins may be deformed.
• After plugging the connector into the terminal, do not subject
the cable to excessive force. Cable damage can result.
• Confirm that the power cords are securely attached so they
will not accidentally be disconnected during operation.
Otherwise, images could disappear during the procedure.
• Never apply excessive force to bend, pull, twist or squeeze
the power cords. Power cord damage can result.
• Never pull connectors or cables with excessive force.
Connector or cable damage can result.
Prepare this instrument, the compatible ancillary equipment (shown in the
“System chart” in the Appendix) and other equipment to be used with this
instrument before each use and refer to the instruction manual for each piece of
equipment.