Chapter 3 Installation and Connection
VIDEO SYSTEM OTV-SI
3.7 Connection to a wall mains outlet
Only connect this instrument to a wall mains outlet. Failure to
do so may cause an electric shock or fire.
Secure the power cord so that it will not be accidentally
disconnected during the procedure. The endoscopic image
will disappear if the power cord is disconnected during use.
Do not allow the power cord to become wet. A wet power
cord may cause an electric shock.
Make sure that the wall mains outlet and isolation
transformer have adequate electrical capacities. Otherwise,
fire or power fluctuations can occur.
Do not bend, pull, twist or squeeze the power cord. an
electric shock, equipment damage or fire can result.
When using non-medical ancillary electrical equipment,
always connect the equipment to a wall mains outlet via an
isolation transformer. Otherwise, an electric shock or fire can
result.
Never use a 3-pin converter plug to connect to a 2-pin wall
mains outlet, as this will prevent proper grounding and could
cause an electric shock.
Never use a table tap. Always connect the power cord
directly to a wall mains outlet. Connecting to a table tap can
result in fire.