Function of the Service Keys and Switches
A. Menu Access Switch
This switch must be open to work with the service software or configure the A/
D converter (e. g., “Trade” configuration). If the menu access switch is closed,
the error message “ACCESS SWITCH LOCKED” is displayed in the service program
and the indicator shows “L” (underload) or “H” (overload). In this case, the A/D
converter cannot be configured. If adjustment is necessary (span, linearity), the
“± 2%” window is active.
B. SBI/BPI Key
To use the Sartocas program (v 1.46 or later) or Psion server (v 4.10 or later).
the Combics indicator must be in BPI (binary processor interface) mode (for
example, to load a modified data record from the A/D converter into its
controller). This setting is valid for the COM1 port.
Note: To perform adjustment
only with the service software, the COM1
interface is set to the xBPI (eXtended Binary Processor Interface) protocol
and the laptop or Psion server connected to that port.
verriegelungs_schalter.eps
closed open
A
sbi_bpi_taster.jpg
B

Table of Contents

Other manuals for Sartorius combics 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 1 and is the answer not in the manual?

Sartorius combics 1 Specifications

General IconGeneral
BrandSartorius
Modelcombics 1
CategoryMonitor
LanguageEnglish

Related product manuals