34
– Press ) until SEt VOR is displayed.
– Release the ) key; after a brief pause, the Combics indicator automatically
returns to the weighing mode.
1-9-9 Clearing the Preload
Note: When using a verified weighing platform, open the cover plate on the left side
of the back of the indicator and move the menu access switch to the right
(“accessible”; see page 14).
– Activate the service mode (see page 17).
– Select weighing platform »WP1«.
– Press the ) key to open the numeric menu.
– Select menu item 1-9-9 (Clear preload) (key sequence: ), then, several times,
k )) and confirm by pressing ) briefly.
– Press and hold the ) key to store the selected menu items and return the
Combics indicator to the weighing mode.
vorl2.eps
0lb_o.eps
display_serv_01.eps
display_wps1_serv.eps
vorl3.eps