Activating the BPI Mode
Turn on the Combics.
In weighing mode, press and hold the SBI/BPI key for approx.
5 seconds (the weight readout goes blank).
Release the switch.
Turn the Combics off and then on again.
The Combics is now in BPI mode.
Important: Peripheral devices that communicate over the SBI (Sartorius Balance Interface)
protocol cannot be operated while the Combics is in BPI mode.
If you press the p key, only InF 30 is displayed.
To return the indicator to the SBI mode, either use the “Close” function in the
Sartocas service software or a Psion server (and then turn the Combics off and
back on again), or select the Menu Reset option (9 - 1 - 1) in the Combics
operating menu.
C. Boot Key
Activate this switch only if an attempt to program the application memory
with the PPLoader program fails (programming routine stops responding).
Procedure for making the PCB bootable:
Disconnect the Combics from the main power.
Press and hold the boot switch (approx. 3 sec.) while reconnecting the Combics
to power.
Release the switch. After a brief period, the PCB is bootable again
(load application software again with the PPLoader program).
boot_taster.jpg
C

Table of Contents

Other manuals for Sartorius combics 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 1 and is the answer not in the manual?

Sartorius combics 1 Specifications

General IconGeneral
BrandSartorius
Modelcombics 1
CategoryMonitor
LanguageEnglish

Related product manuals